More Information

The Electrode Company

Replica Rolex
Replica Watches
Fake Oakleys

The Electrode Company

Pulse Oximeter Sensor Accuracy


Pulse oximeter calibration is achieved during the design phase by carrying out clinical breath down trials with a sensor of a particular specification. The SATs values of volunteers are dropped by using controlled mixtures of oxygen and nitrogen. The R value results from the pulse oximeter are compared with blood samples analysed with a co-oximeter. The results of the breath down trials are commonly referred to as the "R curve". This data is stored in the monitor in software form.

The R curve is a fixed curve for a particular combination of red and infra red wavelengths and enables values of R to be read off as SATs values.



Using a sensor with 664.5nm / 907 nm produces a R value at 1
The R value at 1 intersects the R curve at 2
A corresponding SATs value is read off at 3

If another sensor with different and erroneous wavelengths of, for example, 657nm / 907nm is used simultaneously on the same patient - a different R value at 4 is obtained.
The R value at 4 intersects the inappropriate original R curve at 5
A corresponding but incorrectly low SATs value is read off at 6.

Similarly if erroneous wavelengths of 670nm / 907nm were used the value of R would be lower and an incorrect high SATs value would be obtained.




BACK | NEXT